Mass Flow Controller/Mass Flow Meter With Indicator
Kofloc MODEL D8500 SERIES
Kofloc mass flow controller/meter driven by a 24 VDC power supply has been developed as a successor to the MODEL 8300. The viewpoint change function of the display unit and the pattern setting function are unique to this model, and noise resistance has been improved dramatically. A sister model with a detachable display and setting unit is also available.
Features :
- The high-lift actuator allows this compact model to control a large flow rate.
- Equipped with a display and setting unit, this model can be operated by a 24 VDC power supply
- The RS232C/RS485 communication function and integration function are provided as standard equipment.
- The 14-bit converter permits to display and operation in 4-1/2 digits.
- Control of the flow rate of inflammable gas is possible because the heat generating part of the sensor is not exposed to the gas.
- There are no limitations on the mounting position that may be employed
- In addition to the SV setting, five other patterns can be set.
- Auto zero and auto close functions are also standard.
Standard Specifications :
- Flow range (F.S.) (at N2 calibration conditions) : 50 SCCM-5 SLM
- Applicable gases (dry gas): N2, air, O2, CO2, Ar, H2, He, etc.
- Sensor: a Thermal mass flow sensor
- Valve actuator: Normally-closed solenoid valve actuator *7
- Valve type: Poppet valve *7
- Control System :
- Control/measurement range : 2-100% F.S.
- Response: 0-100% F.S. or more within 2 sec. *1
- bellow 10% F.S. within 4 sec. *
- Accuracy :
- Flow accuracy: ±1.0%F.S. *2
- Repeatability: ±0.75%F.S.
- Pressure :
- Proof pressure: 1000 kPa (G)
- Allowable operating pressure: 500 kPa (G) or less
- Operating differential pressure *7 : 50–300 kPa (G) => 100-300 kPa (G)
- Temperature :
- Allowable operating temperature: 5–45℃
- Temperature characteristics : 0.2% F.S./℃
- Humidity :
- a). Allowable operating humidity: 10–90% (No condensation allowed)
- Leak: He leak rate: 1 × 10-8 Pa•m3/sec. or less *3
- Flow setting method:
- Digital :
- Setting & display unit
- Communications
- Event input selection
- Analog *7 : (1) 0–5 V (2) 4–20 mA (freely selectable)
- Flow rate output:
- Analog : (1) 0-5 V (2) 4-20 mA (interlocked with the above)
- Digital :
- Display :
- Display format: 7-segment 4-digit LED
- Total flow: 12 digits *4
- Mounting direction: Changeable
- Built-in/Separate :Built-in, separate 1 m, separate 3 m, separate 5 m
- Status display LED : OK (within allowable range), ALM (alarm output interlock) OUT1 (event output 1 interlock), OUT2 (event output 2 interlock) SV (set flow), PV (instantaneous flow), TF (total flow) IF (mode setting)
- Other I/O functions :
- Event input : 3 × contact input
- Alarm output: 1 × open collector output, Max. 35 V, 50 mA
- Event output: 2 × open collector output Max. 35 V, 50 mA
- Communications: RS-485, half-duplex, 9600 bps
- Power supply :
- Rating : 24 VDC, current consumption: 300 mA max.
- Allowable supply voltage range: 21.6–26.4 VDC (Ripple: 5% or less)
- Mounting position: Not specified
- Applicable standards: RoHS and EN62326-1: 2006
- Materials of parts in contact with gases: SUS316, SUS316L, SUS430, FKM, PTFE, chloroprene rubber (option)
- Joint: 1/4 SWL, RC1/4, 1/4 VCR
- Weight :
- Built-in: Approx. 1000 g
- Separate (excluding the cable): Approx. 1200 g (*6)